General Information
-Cellstain- Double Staining Kit consists of two staining compounds, Calcein-AM and Propidium Iodide (PI). This kit is utilized for a simultaneous detection of viable cells and dead cells in a cell culture dish by using a fluorescent microscope. Calcein-AM is cell membrane permeable and the generated Calcein by esterase activity in the viable cell emits strong green color fluorescence. Therefore, viable cells can be monitored as green fluorescent-colored cells under the microscope. On the other hand, PI can enter a dead cell through a damaged cell membrane and intercalate DNA of a dead cell. PI emits red color fluorescence in DNA double helix, so dead cells can be monitored as red fluorescent-colored cells.
Kit Contents
Solution A (1 mmol/l Calcein-AM/DMSO) | 50 μl x 4 |
Solution B (1.5 mmol/l PI/purified water) | 300 μl x 1 |
Storage Condition
Store at -20°C
General Protocol
Preparation of a staining solution *
- Before opening vials of Solution A and B, allow them to warm to room temperature.
- 10 μl of Solution A and 15 μl of Solution B are mixed in 5 ml of D-PBS.
- This solution should be used up in one day because of the poor stability of Calcein-AM.
Preparation of cells
- Remove cells from a culture dish by trypsin-EDTA.
- Collect cells by centrifuge (1000 rpm, 3 minutes).
- Discard the supernatant, and add PBS(-) to suspend cells (preferably 105 - 106 cells/ml).
- Repeat step 2 and step 3 several times in order to eliminate esterase activity of the media.
- Mix 100 μl of the staining solution with 200 μl of the cell suspension, and incubate the mixture for 15 minutes at 37°C.
- Observe both living cells stained yellowish-green fluorescent and dead cells stained red fluorescent using an excitation filter at 490±10 nm under fluorescence microscope. In addition, use a 545 nm emission filter to observe only dead cells.
- As the optimal staining conditions differ from cell to cell, we recommend that suitable concentration of PI and Calcein-AM be determined individually.
Precaution
- Since Calcein-AM/DMSO solution is moisture sensitive, please close the cap of Solution A tightly after using.
- PI is suspected to be highly carcinogenic, so careful handling is required. When PI solution contacts skin, immediately wash off using water.
Frequently Asked Questions / Reference
CS01: -Cellstain®- Double Staining Kit
Revised Dec., 11, 2023